Main Features:
1. No consumption of medicament.
2. No film,not affected by surfactant.
3. No micro void,no risk of blocking in sewage.
4. Good applicability, standard two wires structure, 4-20mA output,exchangeable with other film electrode.
5. Easy maintenance,no need to change film.
6. Automatic temperature compensation.
7. No need of pressure balancing.
Parameters |
Description |
Measurement range |
0-0.5/2/5/10/20/50ppm or according to customers’ requirements |
Accuracy |
±0.1% |
Response time |
1-60seconds |
Temperature |
5-50centigrade |
Output |
4-20mA |
Supply voltage |
12-24V |
velocity |
20-40l/min |
Materials |
POM、PTFE、Gold、Ag selective ion exchange ` |
Working life of electrolyte |
1 year |
Interface |
Standard two/four wires interface |
Dimension |
φ25*140mm |
Calibration |
1 slope calibrate, calibrate zero |
Installation method |
Flow cell installation |